Power
supply
The instrument is powered by
12V±0.3V DC and draws
2.1A. Current is supplied with
an IP67 waterproof Binder series
768 connector. Power on is indicated
by an LED on the rear side of
the instrument.
Communication
interface
Measurement results are transmitted
via RS232 and the series communication
interface is bidirectional for
reception of control instructions.
The RS232 interface uses 19200
baud, 8 data bits, one stop
bit and no parity. Measurement
results are transmitted from
the instrument every 125ms using
a proprietary protocol.
A 9 pin IP67 waterproof LTW
D-sub connector is used, shown
at the lower right side in the
image. |
|
Reset
The instrument is provided
with a reset button, visible
on the lower left side of the
image below. Software controlled
reset via RS232 is also available.
 |
| Rear side |
Ports
not intended for customer use
A lid, shown at the centre
of the image above, prohibits
access to further ports that
are not intended for normal
customer use. The lid should
not be removed.
The instrument is calibrated
at production for optimal performance
at ordinary use. It may be recalibrated
using a switch behind the lid,
but if incorrectly done it may
reversibly affect performance
negatively and this is strongly
advised against. |